Dilatometer DIL 803

Measurement of two sample at the same time
   

 

Zoom DIL801
   
   
 

The DIL 803 is a two-sample dilatometer designed for high accuracies with high sample throughput.
Measurements may be carried out under vacuum and inert gas and within the temperature range
of -160°C to 2400°C.

The measuring head is thermally stabilized and insensitive to shocks. You can use easily interchangeable measuring systems made of fused silica, Al2O3, sapphire, graphite and tungsten.
The temperature measurement is carried out with thermocouples (type K, S, B, C) and radiation pyrometers.

The inductive sensor (L.V.D.T.) has a resolution of 10nm.
Various certified reference materials for the calibration of the equipment are on offer.
The entire electronics operates digitally and with an own processor system to control the furnaces, the measuring data recording, the gas supply, the vacuum unit and safety devices.

The professional 32-Bit software WinTA 9.0 runs under Windows®98, ME, 2000, NT4.0, XP.
An extensive software accurately and efficiently deals with both routine tasks and comprehensive evaluations. Export functions as tables in ASCII- and Excel format and a graphic file output in the formats BMP, JPG, TIF, EMF etc. enable a cross-platform work.

 

   
   
   
 
 
 
 
   
Features DIL 803
   
Sample length: 0 to 50 mm
Sample diameter: max. 7 or 10 mm
Sample holder material: fused silica, Al2O3, sapphire, graphite, tungsten
Change of length: 4mm
Resolution l °C: 10nm, 0.05°C
Accuracy in : 0,03 x 10-6 K-1
Atmosphere: vacuum, inert gas, air
Operation mode: horizontal
Temperature range: -160°C to 2400°C according to furnace type
Contact force: 0,02 to 1N, adjustable